The Commonest Cause of a Pause

Report:

Sinus tachycardia 100/min 1

Atrial-sensing ventricular pacemaker rhythm 1

Blocked SVEB 6

Left atrial abnormality (LAA) 1

Right bundle branch block 1

Comment:

There is a subtle but definite change in the T wave of the last paced beat before the pause: it is more peaked in lead 2 rhythm strip. The pacemaker may have failed to sense this P wave, or the latter may be within the pacemaker’s PVARP.

The sinus escape P wave just beats the pacemaker to the ventricles, its PR interval being shorter than the programmed (sensing) AV interval. This is not obvious from the trace itself; the difference may be in milliseconds and we do not know which lead the pacemaker electrode is “looking” at.

Below (Fig 100a) is another example, from a different patient; the T wave is dimpled in the V1 rhythm strip. This time the escape is paced atrial P wave, followed by a native atrial wave P’ in the next cycle and then resumed sinus rhythm.

As Marriott said: “The commonest causes of pauses are blocked APCs”53.

Fig 100a.

Fig 101. 88 year old man several days after aortic valve replacement.

If you have any suggestions for or feedback on this report, please let us know.